High-function TOC-620

TThe TOC-620 is an automatic TOC analyzer which quickly and accurately measures the total organic carbon content of water from rivers, lakes, marshes, sea areas, factory wastewater and other sources, in either online or offline mode.
This system employs Toray Engineering's unique low-temperature sealed combustion system to ensure long-term, stable measurement. It is designed, based on the track record of the previous TOC-610 model, for greater compactness assuming use of the dilution function (standard) and stream switching function.

Features

Principle of Measurement

After removing inorganic carbon from the sample water in an aeration cylinder, and measuring it in the measuring section, the sample is injected into a reaction tube equipped with catalyst, and converted to carbon dioxide through sealed combustion oxidation. Concentration of the converted carbon dioxide is measured with an infrared analyzer, and the TOC value is obtained from the calibration curve for carbon dioxide and TOC stored beforehand in the signal processing unit.

Specifications

Measurement range 0-100 / 1000 mg/l (High concentration) 0-5-50 mg/l (Low concentration)
Dilution function 2-20 times
Stream switching function 2-6 streams *1
Measurement cycle From 4 min
Measurement method Inorganic carbon removal: Automatic intermittent HCl acidic air aeration method
Total carbon measurement: Automatic intermittent sealed combustion infrared analysis method
Calibration method Span calibration: Manual and automatic calibration (Option)
Zero calibration: Manual (With auto zero function)
Conversion function Conversion from TOC value to COD value and display
Washing Automatic washing of injection tube and combustion section; Alternate injection of sample and washing liquid (Washing liquid contained in the unit)
Range setting Manual switching
Construction Standalone console, with casters
Dimensions 602W × 500D × 1570H mm
Weight Approx. 110 kg
Standard liquid Potassium hydrogen phthalate aqueous solution
Repeatability Within ±2 % of full scale Within ±3 % of full scale
(5mg/l range)
Zero stability Within ±2 % of full scale/day Within ±3 % of full scale
(5mg/l range)
Span stability Within ±2 % of full scale/day Within ±3 % of full scale
(5mg/l range)
Ambient temperature variation Within ±2 % of full scale/5°C Within ±3 % of full scale/5°C
(5mg/l range)
Measurement value output External output: DC 4-20 mA
RS232C (Option)
Range ID signal no-voltage contact: AC 125 V, 1A
Error alarm indication LED display, Closed contact output in case of error (AC125, 2A)
Alarms Heater temperature control
Upper/Lower limits for measured value
Out of carrier gas (Option)
No sample (Option)
Out of reagent (Option)
Out of dilution water (Option)
Out of washing water (Option)
Sample conditions Online measurement: 0.02-0.1 MPa, 500 ml/min or higher
Offline measurement: Normal pressure, 200-500 ml/3 times
Reagents / Utilities Washing liquid: Distilled water
Reagents: HCl 2 mol/l, 2.0 l/month
Carrier gas: Nitrogen gas *2, Pressure 0.3 MPa, Consumption 7 Nm3/month
Power AC 100 V±1 0V 50/60 Hz Approx. 350 VA
Set up conditions Indoors
3-40 °C during operation (Temperature gradient 10 °C /hour or less)
45 %-85 %RH (No condensation)

Specifications are subject to change due to product improvements.
*1 The stream switching function is an option. Please designate the necessary streams before determining specifications.
*2 The carrier gas can be changed to air specifications. However, please consult with the sales supervisor when making that change.

High-function TOC-620
Water Quality Analyzers
Space-saving, Low-cost TOC-100
High-function TOC-620
For Pure Water and Recycled Water Applications TOC-780
Automatic TOC Analyzer TOC-106
Total Organic Carbon (TOC) Analyzer for Laboratory TOC-150
Automatic Total Oxygen Demand Analyzer TOD-810C
Automatic UV Analyzer UVT-300
Inquiry
Please use this form to request information or make an inquiry regarding High-function TOC-620.
To an inquiry form
TOP > Measurement & Control Technologies > Water Quality Analyzers